Chapter 4 ______________________________________________________________ Maintenance

VAISALA _______________________________________________________________________ 35

Connections and Selecting the Calibrator Function

1.

To select the calibrator function from the HMI41, press the

ON/OFF

button until you can see some text on the display. Then release the

ON/OFF

button and within 1 ... 2 seconds, press both

ENTER

and

MODE

buttons until the following display appears:

2.

After a few seconds, the display changes to show the following:

set

°C

If the basic settings (display units, automatic power off function,
display quantities, and pressure) have to be changed, please refer
to the HMI41 Operating Manual. Otherwise, press

ENTER

repeatedly until the following display appears:

set

To calibrate the HMT100, select number 3 with buttons ▲
(number up) and ▼ (number down) and then press

ENTER

.

3.

Next the baud rate appears on the display. Use the baud rate 19200
(= 19.2 on the HMI41 display) with the HMT100 series. If the
baud rate on the HMI41 display is not correct, change it with
buttons ▲ and ▼. When the baud rate is correct, press

ENTER

, and

the serial communications settings display appears. The correct
settings for HMT100 series are: N, 8, 1. If needed, change the
settings with buttons ▲ and ▼ until they are correct, press

ENTER

,

and then

ON/OFF

.

These setting are stored in the HMI41 memory; when the HMI41 is
turned on again, it will automatically wake up as a calibrator for digital
transmitters with these serial line settings. After making these settings,
continue with the adjustment and follow the directions of the chosen
adjustment method.