Two-point adjustment using a calibrator, Licl-nacl adjustment – Vaisala HMT100 User Manual

Page 32

background image

User's Guide ______________________________________________________________________

32 __________________________________________________________________ M210701EN-D

Two-Point Adjustment Using a Calibrator

Note that the difference between the two reference humidities must be at
least 50 %. MI70 indicator is used now only as a terminal for visualizing
and setting the transmitter's RH reading.

Follow steps 1 to 7 on page 30 and continue as follows:

8.

Remove the filter from the transmitter's probe and insert the
probe head into the lower humidity reference condition.

9.

Press

ADJUST

to continue adjusting.

10. Choose

2-point adjustment

from the MI70 adjustment menu and

press

SELECT

.

11. Press

READY

when the reading has stabilized in the first reference

condition (may take 30 minutes or more). You can follow the
stabilization from the

GRAPH

display.

12. Enter the correct reference value in the first condition with the

arrow buttons. Press

OK

.

13. Remove the probe from the first reference condition and insert

the probe head into the higher humidity reference condition.

14. Press

READY

when the reading has stabilized in the second

reference condition (may take 30 minutes or more). You can
follow the stabilization from the

GRAPH

display.

15. Enter the correct reference value in the second condition with

the arrow buttons. Press

OK

.

16. Confirm the adjustment by pressing

YES

(by pressing

NO

you

return to adjustment mode display and no changes are made). If
the difference between the two reference conditions is less than
50 %RH, adjustment cannot be done.

17. The adjustment is done. Press

BACK

and

EXIT

to return to the

basic display.

18. Turn off the MI70 and detach the connection cable.

LiCl-NaCl Adjustment

This adjustment is done using relative humidity references 11.3 % RH
(LiCl) and 75.5 % RH (NaCl).

Follow steps 1 to 7 on page 30 and continue as follows:

8.

Remove the filter from the transmitter's probe and insert the
probe head into the LiCl salt chamber.

9.

Press

ADJUST

to continue adjusting.