Mi70 indicator, Indicator general, Mi70 indicator indicator general – Vaisala HM70 User Manual
Page 77
Chapter 14 ____________________________________________________ Technical Specifications
VAISALA _______________________________________________________________________ 75
Housing material
ABS/ PC blend
Probe material
Stainless steel (AIS316L)
Probe cable length
(between indicator and the probe handle)
1.9 m
Probe cable length of HMP77
(from handle to the root of the probe)
5.0 m
Probe diameter
12 mm (0.47 inch)
Weight
HMP75
250
g
HMP76
350
g
HMP77
500
g
MI70 Indicator
Indicator General
Operating temperature range
-10 ... + 40 °C
Operating humidity
0 ... 100 %RH , non-condensing
Menu languages
English, Finnish, French, German,
Spanish, Swedish, Chinese,
Russian, Japanese
Display
LCD with backlight
Graphic trend display of any
quantity
Character height up to 16mm
Probe inputs
1 or 2
Power supply
Rechargeable NIMH battery pack
with AC-POWER or 4 x AA size
alkalines, type IEC LR6
Analog output
0 ... 1 VDC
Output resolution
0.6 mV
Accuracy
0.2 % full scale
Temperature dependence
0.002 %/°C full scale
Minimum load resistor
10 kΩ to ground
Data interface
RS232C (EIA-232)
Data logging capacity
900 ... 2700 real time data points,
Logging interval
1 s to 12 h
Logging duration
1 min ... memory full