User's Guide ______________________________________________________________________

54 __________________________________________________________________ M210297EN-E

9.

Remove the filter from the transmitter's probe and insert the
probe into the reference condition.

10. Press

ADJUST

to continue adjusting

.

11. Select

1-point adjustment

, press

SELECT.

12. Press

READY

when the reading is stabilized in the reference

condition (takes about 30 minutes or more)

.

You can follow the

stabilization from the

GRAPH

display.

13. Enter correct reference value by using the arrow buttons. Press

OK

.

14. Confirm by pressing

YES.

15. Adjustment is done. Press

BACK

and

EXIT

to return to the

basic display.

16. Switch off the MI70 and detach the calibration cable.

2-Point Adjustment by Using a
Calibrator

Note that the difference between the two reference humidities shall be
at least 50 %. MI70 indicator is used now in communication.

If you use humidity calibrator HMK15, please use the adapter fitting
(13.5 mm) on the measurement hole if you calibrate probes of
HMT334, HMT335, HMT337 and HMT338.

Follow the steps from 1 to 8 on page 52 and continue as follows:

9.

Remove the filter from the transmitter's probe and insert the
probe into lower reference humidity.

10. Press

ADJUST

to continue adjusting

.

11. Select

2-point adjustment

, press

SELECT.

12. Press

READY

when the reading is stabilized in the first

reference condition (takes about 30 minutes or more)

.

13. Enter correct reference value by using the arrow buttons. Press

OK

.

14. Remove the probe and insert the probe into higher reference

humidity. You can follow the stabilization from the

GRAPH

display (takes about 30 minutes or more).

15. Press

READY

when the reading is stabilized in the second

reference condition

.