Pressure Setting

When measuring in pressurized environments, the actual process
pressure value must be given to the HM70. The pressure can be given
in the following units:

hPa: Absolute pressure given in the unit of hPa.

barg

:

Gauge pressure given in the unit of bar. Indicates the pressure
difference between normal atmospheric pressure and the actual
process pressure.

bara

:

Absolute process pressure given in the unit of bar.

psia

:

Absolute pressure given in the unit of psi.

Follow the instructions below:

1.

Open the

MENU;

press

OPEN

2.

Select

ENVIRONMENT

with

and press

.

3.

Press

UNIT

to change the pressure unit. Default unit is hPa.

4.

Press

SET

to set the ambient pressure value.

5.

Set the pressure value by using buttons

.

Press

+/-

to

change the sign of the pressure value. Press

OK

to save the

value.

6.

Press

EXIT

to return to the basic display.