Chapter 9 ______________________________________ Calibration and Adjustment of Transmitters

VAISALA _______________________________________________________________________ 53

From now on the adjustment is carried out by using the MI70
indicator.

7.

Press

OK

to start adjustment.

8.

Select

RH

or

T,

press

SELECT.

9.

If RH selected, check the environment settings, if needed.
Otherwise press

NO.

Field Checking and Adjustment by
Using a Calibrated Reference Probe

Follow the steps from 1 to 8 on page 52 and continue as follows:

9.

Check that the probes are located in equal conditions and wait
until the readings are stabilized (can take 30 minutes or more). If
you are near the probes, do not breathe in their direction.

10. Press

ADJUST

to continue adjusting

.

11. Press

To same asRH

Ii/I

, press

SELECT.

(MI70 recognizes always

the port to which the HMP70-series probe is connected).

12. Confirm by pressing

YES.

13. Adjustment is done. Press

BACK

and

EXIT

to return to the

basic display.

14. Switch off the MI70 and detach the calibration cable.

1-Point Adjustment by Using a
Calibrator

When adjusting a transmitter only in one reference condition, please
take care that the reference condition represents the measuring
environment. MI70 indicator is used now in communication.

If you use humidity calibrator HMK15, please use the adapter fitting
(13.5 mm) on the measurement hole if you calibrate probes of
HMT334, HMT335, HMT337 and HMT338.

Follow the steps from 1 to 8 on page 52 and continue as follows: