Point adjustment – Vaisala HM70 User Manual

Page 66

background image

User's Guide ______________________________________________________________________

64 __________________________________________________________________ M210297EN-E

1.

Take out the screw covering the calibration button (in the probe
handle).

2.

Press the calibration button. For pressing, use tool with thin and
sharp point, such as small screwdriver. When pressing the
button, indicator turns to adjusting mode.

3.

Select

T

, press

SELECT

.

4.

Now the adjustment mode is on, press

ADJUST

to select the

adjustment method:

1-point adjustment

or

2-point adjustment

.

1-Point Adjustment

Follow the steps from 1 to 4 and continue as follows:

5.

Select

1-point adjustment,

press

SELECT.

6.

Set the probe to a reference temperature. You can follow the
stabilization from the

GRAPH

display. Press

READY

when the

reading is stabilized in the reference.

7.

Give the reference temperature value by using the arrow
buttons. Press

OK.

8.

Confirm the adjustment, press

YES

. (By pressing

NO

you

return to adjustment mode display and no changes are made).

9.

Calibration is carried out. Press

BACK

to exit the adjustment

mode and

EXIT

to return to the basic display.