Licl-nacl adjustment – Vaisala HM70 User Manual

Page 47

background image

Chapter 9 ______________________________________ Calibration and Adjustment of Transmitters

VAISALA _______________________________________________________________________ 45

LiCl-NaCl Adjustment

This adjustment is done using relative humidity references 11.3 % RH
(LiCl) and 75.5 % RH (NaCl).

Follow steps 1 to 7 on page 42 and continue as follows:

8.

Remove the filter from the transmitter's probe and insert the
probe into the LiCl salt chamber.

9.

Press

ADJUST

to continue adjusting.

10. Choose

LiCl-NaCl autom.

from the MI70 adjustment menu, press

SELEC

T. Press OK to accept the note telling about references.

11. Press

READY

when the reading has stabilized in the LiCl salt

chamber (may take 30 minutes or more). You can follow the
stabilization from the

GRAPH

display.

12. Remove probe from the LiCl salt chamber and insert the probe

into the NaCl salt chamber.

13. Press

READY

when the reading has stabilized in the NaCl salt

chamber (may take 30 minutes or more). You can follow the
stabilization from the

GRAPH

display.

14. Confirm the adjustment by pressing

YES

(by pressing

NO

you

return to adjustment mode display and no changes are made).

15. The adjustment is done. Press

BACK

and

EXIT

to return to the

basic display.

16. Turn off the MI70 and detach the connection cable.

Calibration and Adjustment of HMP230/240
Transmitters and HMM210 Modules

You are able to check and adjust only humidity channel, temperature
channel can only be checked. Calibrate your transmitter/module
against a calibrated reference probe of HM70 or against a calibrator's
reference humidity by using MI70 indicator in communication.
Follow the first 12 steps and continue according to the chosen
calibration method.

1.

In case you have a transmitter provided with heated probe, carry
out these preparations before calibration.