User's Guide ______________________________________________________________________

48 __________________________________________________________________ M210297EN-E

14. Press

ADJUST

15. Select

1-point adjustment

. press

SELECT

16. Wait until the reading is stabilized (takes about 30 minutes).

You can follow the stabilization from the

GRAPH

display.

17. Press

READY

when the value is stabilized in the reference

humidity.

18. Enter the correct humidity value by using the arrow buttons.

Press

OK

.

19. Confirm by pressing

YES

20. Press

BACK,

EXIT,

EXIT

to return to the basic display.

2-Point Adjustment by Using a
Calibrator

You can use Vaisala humidity calibrator HMK15 in calibration and
adjustment. Note that the difference between the two reference
humidities shall be at least 50 %. MI70 indicator is working as a
communicator in the calibration procedure.

Follow the steps from 1 to 12 and continue as follows:

13. Insert the transmitter probe to the first reference humidity.

14. Press

ADJUST

15. Select

2-point adjustment

. press

SELECT

16. Wait until the reading is stabilized (takes about 30 minutes).

You can follow the stabilization from the

GRAPH

display.

17. Press

READY

when the value is stabilized in the first reference

humidity.

18. Enter the first reference humidity value by using the arrow

buttons. Press

OK

.

19. Insert the transmitter probe to the second reference humidity.

20. Wait until the reading is stabilized (takes about 30 minutes or

more). You can follow the stabilization from the

GRAPH

display.

21. Press

READY

when the value is stabilized in the second

reference humidity.

22. Enter the second reference humidity value by using the arrow

buttons. Press

OK

.