User's Guide ______________________________________________________________________

50 __________________________________________________________________ M210297EN-E

Field Checking and Adjustment by
Using a Calibrated Reference Probe

Follow the steps from 1 to 8 and continue as follows:

9. Check that the probes are located in equal conditions and wait until

the readings are stabilized (can take 30 minutes or more). If you
are near the probes, do not breathe in their direction.

10. Press

ADJUST

to continue adjusting

.

11. Press

To same asRH

Ii/I

., press

SELECT.

(MI70 recognizes always the

port to which the HMP70-series probe is connected).

12. Confirm by pressing

YES.

13. Adjustment is done. Press

BACK

and

EXIT

to return to the

basic display.

14. Replace the

CAL.DIS

jumper.

15. Switch off the MI70 and detach the calibration cable from MI70

and from the transmitter.

1-Point Adjustment by Using a
Calibrator

When adjusting a transmitter only in one reference condition, please
take care that the reference condition represents well the measuring
environment.

If you use humidity calibrator HMK15, please use the adapter fitting
(13.5 mm) on the measurement hole if you calibrate probes of
HMT324, HMT325, HMT327 and HMT328.

Follow the steps from 1 to 8 and continue as follows:

9.

Remove the filter from the transmitter's probe and insert the
probe into the reference condition.

10. Press

ADJUST

to continue adjusting

.

11. Select

1-point adjustment

, press

SELECT.

12. Press

READY

when the reading is stabilized in the reference

condition (takes about 30 minutes or more)

.

You can follow the

stabilization from the

GRAPH

display.