beautypg.com

Accuracies of calculated variables, Accuracy of dewpoint temperature °c, Accuracy of mixing ratio g/kg – Vaisala HM70 User Manual

Page 75

background image

Chapter 14 ____________________________________________________ Technical Specifications

VAISALA _______________________________________________________________________ 73

Wet bulb temperature

0 ... +100 ºC, 32 ... +212 ºF

Enthalpy

0 ... 160 kJ/g

Water vapor pressure

0 ... 1000 hPa

Accuracies of Calculated Variables

Accuracies of the calculated variables depend on the calibration
accuracy of the humidity and temperature sensors; here the accuracies
are given for

±

2 %RH and ±0.2 °C

.

Accuracy of Dewpoint Temperature °C

Relative

humidity

Temp.

10

20 30 40 50 60 70 80 90 100

-40

1.86

1.03 0.76 0.63 0.55 0.50 0.46 0.43

-20

2.18

1.19 0.88 0.72 0.62 0.56 0.51 0.48

0

2.51

1.37 1.00 0.81 0.70 0.63 0.57 0.53 0.50 0.48

20

2.87

1.56 1.13 0.92 0.79 0.70 0.64 0.59 0.55 0.53

40

3.24

1.76 1.27 1.03 0.88 0.78 0.71 0.65 0.61 0.58

60

3.60

1.96 1.42 1.14 0.97 0.86 0.78 0.72 0.67 0.64

80

4.01

2.18 1.58 1.27 1.08 0.95 0.86 0.79 0.74 0.70

100

4.42

2.41 1.74 1.40 1.19 1.05 0.95 0.87 0.81 0.76

120

4.86

2.66 1.92 1.54 1.31 1.16 1.04 0.96 0.89 0.84

140

5.31

2.91 2.10 1.69 1.44 1.26 1.14 1.05 0.97 0.91

160

5.80

3.18 2.30 1.85 1.57 1.38 1.24 1.14 1.06 0.99

Accuracy of Mixing Ratio g/kg

(ambient pressure 1013 mbar)

Relative

humidity

Temp.

10 20

30 40 50 60 70 80 90 100

-40

0.003 0.003

0.003 0.003 0.003 0.004 0.004 0.004

-20

0.017 0.018

0.019 0.021 0.022 0.023 0.025 0.026

0 0.08

0.09

0.09

0.10

0.10 0.11 0.11 0.12 0.13 0.13

20 0.31 0.33

0.35

0.37

0.39 0.41 0.43 0.45 0.47 0.49

40 0.97 1.03

1.10

1.17

1.24 1.31 1.38 1.46 1.54 1.62

60 2.68 2.91

3.16

3.43

3.72 4.04 4.38 4.75 5.15 5.58

80 6.73 7.73

8.92

10.34

12.05 14.14 16.71 19.92 24.01 29.29

100

16.26 21.34

28.89 40.75 60.86 98.85 183.66

438.56

120

40.83

74.66 172.36 —