beautypg.com

Hmi41 indicator, General – Vaisala HM44 User Manual

Page 47

background image

Chapter 12 ____________________________________________________________ Technical data

VAISALA __________________________________________________________________________

45

°C

0.6

0.4

0.2

0

-0.2

-0.4

-0.6

-20

-10

0

10

20

30

40

50

60 °C

Temperature sensor

Pt1000 IEC 751 Class B

General

Cable length:
HMP44

0.3 m

HMP44L

2.7 m

Probe diameter

12 mm

Probe length

69 mm

Operating temperature range of
electronics

-40...+60

°C

Housing material

ABS plastic

Housing classification

IP65 (NEMA 4)

Sensor protection

membrane filter 17039HM

Weight:

HMP44

50 g

HMP44L

120 g

HMI41 indicator

Maximum measurement
error caused by the indicator at 20

°C:

humidity

+0.1 %RH

temperature

+0.1

°C

Calculated variables

dewpoint temperature, absolute humidity,
wet bulb temperature, mixing ratio

Resolution

0.1 %RH; 0.1

°C

Power supply

4 batteries, type AA (IEC LR6)