Chapter 9, Maintenance, Replacing the humicap180 sensor – Vaisala HM44 User Manual

Page 41: Replacing the filter, Replacing the humicap, 180 sensor

background image

Chapter 9 _______________________________________________________________Maintenance

VAISALA __________________________________________________________________________

39

CHAPTER 9

MAINTENANCE

Replacing the HUMICAP

180 sensor

Remove damaged sensor and insert a new one. Handle the sensor from the plastic frame; do
not touch the sensor plate. Calibrate the probe (with a new sensor) with the calibrator
HMK15. The probe can be sent to Vaisala service for sensor change.

Replacing the filter

The combination of a grid (of the probe head) and the membrane filter can be changed if
needed (order code 17039HM).