COLLECTING DATA

The HMI41 indicator can be used for collecting the measurement data. The data is
stored in the indicator's non-volatile memory; it is not lost when the indicator is
turned off. Please, note also that the automatic power-off function is not active
during data collecting even if previously selected. When data collecting has
ended, the automatic power-off function becomes active again.

Data collecting modes

The following three modes are used when collecting data:

-

REC AUTO

: automatic data collecting

-

REC CATCH

: manual data collecting

-

REC READ

: reading the measurement results

Entering the data collecting modes

Turn the device ON (by using ON/OFF button) and immediately (within 1..2 seconds) after
turning on press immediately the button HOLD for at least 5 seconds. The software version
and the probe type indication appear for a second, after which the text

REC AUTO

appears on

the display; then release the HOLD button.


REC AUTO

is the main display of the data collecting mode. With the button MODE you can

enter the

REC CATCH

mode (manual data collecting, see page 24) and by pressing MODE

again, the

REC READ

mode (reading the measurement results, see page 25).