beautypg.com

Numbering the probes, Feeding the offset correction, Feeding the gain correction – Vaisala HM44 User Manual

Page 10

background image

User's Guide _______________________________________________________________________

_________________________________________________________________________________

8

Numbering the probes

The first number of the second line of display indicates the currently selected probe (1-10).
Connect a probe and change the number on the display by pressing ENTER and
simultaneously buttons

▲ or ▼. Remember to write the probe number (1-10) in the product

label to be able to identify the probes.

Feeding the offset correction

1. Press the button ENTER continuously for about 5 seconds until a text similar

to the following appears on the display (numbers on the second line are
blinking):

RH

set

2. Feed the individual offset correction (see the calibration certificate) in the

HMI41 memory with buttons

▲ and ▼. Each time the button is pressed, the

numbers change 0.1 %RH. If the button is pressed continuously, the numbers
change in a faster rate.

3. When the reading is correct, press ENTER and the display changes to show

the gain correction.

Feeding the gain correction

RH

set

4. Numbers on the second line are blinking; you can now enter the individual

gain correction in the HMI41 memory. The numbers are changed with
buttons

▲and ▼. Each time the button is pressed, numbers change 0.001. If

the button is pressed continuously, the numbers change in a faster rate.

NOTE

If the correction given in the calibration certificate has only two
decimals (e.g. 1.03), enter a zero in place of the third decimal (e.g.
1.030).