User's Guide _______________________________________________________________________
_________________________________________________________________________________
12
Setting the probe type
In cases where an indicator (or a probe) is without an ID-marking, the following probe
numbers have to be selected:
Probe
Probe setting
HMP41
1
HMP42
2
HMP44
1...10
HMP45
1
HMP46
2
• change the setting with buttons ▲ and ▼
With the probe HMP44 there is no need to change the setting.
• press ENTER; the display now changes to show the text START. Change this
setting only if there is no ID-marking on the indicator label. Otherwise do not
change this setting.
Select the function
In cases where an indicator (or a probe) is without an ID-marking on the label the following
function selection is made:
Probe
Setting
HMP41
HMP42
HMP45
HMP46
Start 1
HMP44
Start 5
• change the setting with buttons ▲ and ▼
• press ENTER three times; the display now changes to show the following
settings (do not change them):
baud
4.8
seri
E.7.1
calib
def (*
(*
when entering the setting calib, the value is always ‘no’; the factory settings are
restored by selecting ‘def’. After this, the previously entered adjustment coefficients
are no longer valid.