User's Guide _______________________________________________________________________

_________________________________________________________________________________

12

Setting the probe type

In cases where an indicator (or a probe) is without an ID-marking, the following probe
numbers have to be selected:

Probe

Probe setting

HMP41

1

HMP42

2

HMP44

1...10

HMP45

1

HMP46

2

• change the setting with buttons ▲ and ▼
With the probe HMP44 there is no need to change the setting.

• press ENTER; the display now changes to show the text START. Change this

setting only if there is no ID-marking on the indicator label. Otherwise do not
change this setting.

Select the function

In cases where an indicator (or a probe) is without an ID-marking on the label the following
function selection is made:

Probe

Setting

HMP41
HMP42
HMP45
HMP46

Start 1

HMP44

Start 5

• change the setting with buttons ▲ and ▼

• press ENTER three times; the display now changes to show the following

settings (do not change them):

baud

4.8

seri

E.7.1

calib

def (*

(*

when entering the setting calib, the value is always ‘no’; the factory settings are

restored by selecting ‘def’. After this, the previously entered adjustment coefficients
are no longer valid.