4.9 CALIBRATION SCREEN
CALIBRATION
To re-zero the device, disconnect the electropneumatic Header H1 so
that room pressures are not applied to the pressure sensor and lightly
press (or tap) the “Zero” button. If a suffi ciently accurate Pressure
Calibrator is available, apply a steady full-scale pressure signal to the
“+” on the header fi tting and press (or tap) the “Span” button. Reconnect
the room pressure tube and calibration is complete. Calibration must
be within ±5% of original calibration for this to occur otherwise an error
message will occur.
The original factory pressure calibration can be restored, if desired, by
pushing the “Restore Factory Setting” button.
See also other documents in the category Setra System Sensors:
- Model ASM (8 pages)
- Model 204 (4 pages)
- Model 209 (4 pages)
- Model 227 (8 pages)
- 264 (4 pages)
- 269 (4 pages)
- Model 280E (4 pages)
- Model 370 (48 pages)
- Model 526 (7 pages)
- Model 540 (9 pages)
- Model 595 (8 pages)
- Model 869 (28 pages)
- Model 321 (4 pages)
- Model SRH (12 pages)
- Model Datum 2000 (31 pages)
- Model 217 (8 pages)
- Model 230 (12 pages)
- Model 242901-06 (1 page)
- Model 265 (4 pages)
- Model 270 (5 pages)
- Model 299 (4 pages)
- Model 470 (35 pages)
- Model 524 (6 pages)
- Model 546 (8 pages)
- Model 730 (16 pages)
- GCT-225 (8 pages)
- Model 141 (4 pages)
- Model 205-2 (4 pages)
- Model 223 (8 pages)
- Model 231 (12 pages)
- Model 256 (8 pages)
- Model 276 (2 pages)
- Model 3100 (8 pages)
- Model 516 (6 pages)
- Model 550 (9 pages)
- Model 760 (18 pages)
- Model LD330 (4 pages)
- Model SRMD (8 pages)
- Model 201 (4 pages)
- Setra Model 206 (4 pages)
- Model 224 (8 pages)
- Model 231RS (16 pages)
- 260 (2 pages)
- 267MR (14 pages)
- Model 278 (2 pages)