6.9 SYS STATUS

Definition

This function displays the software version of the instrument, the RS232 configuration * (baud rate,
parity, bits and stop bit), the network address *, any user defined description for the instrument *
and the serial number of the instrument.

This information is for display only. Some parameters such as the network address can be
changed with the HW4 software, after connecting the instrument to a PC.

* except HygroLab 1


Procedure

When entering the function, the software version number is displayed first. Use the ▼ or ▲ key to
display the other data.

7

Error and status messages

The following is a list of coded messages (101, etc.) that the HygroLab may show on the bottom
line of the LC display or via RS232 / RS485.

Errors:

101

checksum error

The checksum test did not pass during RS-communication.

102

bad command

An unknown command was received

103

disallowed
command

A command was received that is reserved for production and
service

104

unknown probe
input

Reference was made to a non-existing analog or digital probe
input

105

argument error

Error in the number of the arguments in the command or in
the value of one of the arguments

106

HygroClip
communication error

The HygroClip probe does not answer or is not connected

107

calibration error

The difference between the probe reading and the calibration
point is larger than the maximum allowed by the INI
command.

108

calibration error
(overflow)

Internal probe error (or the difference between the probe
reading and the calibration value is too large)

110

unknown reference
probe

The reference probe is not connected or the reference probe
input does not exist

111

Temperature error

During humidity calibration, temperature should be within the
limits of 0 and 80

°C (32 and 176°F)

Warnings:

120

no adjustment

Calibration

121

No HygroClip probe
is connected

Status:

130

the probe was
adjusted

Calibration