Calculate (hygrolab 2 and 3), In-e-hylab-v4_11, 1 calculate (hygrolab 2 and 3) – ROTRONIC HygroLab User Manual

Page 17

background image

IN-E-HyLab-V4_11

Rotronic AG
Bassersdorf, Switzerland

Document code

Unit

Instruction Manual

Document Type

HygroLab bench top indicator version 4:
instruction manual

Document title

Page 17 of 51

© 2006; Rotronic AG IN-E-HyLab-V4_11

6.1 CALCULATE (HygroLab 2 and 3)


Definition

This function is used to select or view the humidity parameter that is calculated by the instrument.
The calculated parameter is selected individually for each probe connected to the instrument

1)

.

Prior to entering this function, select on the display the probe to program.

1) Does not apply to third-party analog probes.

The unit system (Metric or English) can be changed with the keypad (see Functions, Settings) or
with the optional HW4 software, when the HygroLab is connected to a PC.

Selections

Parameter

Metric

English

Lab 1

Lab 2

Lab 3

Dew Point / Frost Point

1)

°C

°F

N/A

x x

Wet Bulb Temperature

2)

°C

°F

N/A

x x

Enthalpy

2)

J/g

BTU/lb

N/A

X X

Vapor Concentration

g/m

3

gr/cuft

N/A

X X

Specific Humidity

2)

g/kg

gr/lb

N/A X X

Mixing Ratio

2)

g/kg

gr/lb

N/A

X X

Vapor Concentration
at Saturation

g/m

3

gr/cuft

N/A

X X

Partial Pressure
of Water Vapor

hPa

PSI or In Hg

N/A

X X

Saturation Pressure
of Water Vapor

hPa

PSI or In Hg

N/A

X X

1) The standard factory setting is frost point for values below freezing. This setting can be changed to dew
point with the optional HW4 software
2)

Calculation of this parameter requires barometric pressure as an input. The fixed value used for barometric

pressure can be changed either directly from the keypad - see MENU, SETTINGS - or with the optional HW4
software (all models). The HygroLab 2 or 3 can also be configured to accept the input from an analog
pressure probe (variable pressure value).