Introduction, Installation – ROTRONIC ATEX User Manual
Page 11
11
1. Introduction
The intrinsically safe transmitters HygroClip IC–EX, ID–EX, IE–EX and IW–EX are instruments for the combined measurement
of humidity and temperature in hazardous areas. The transmitter fulfills the European Standards EN 50014, EN 50020, EN50284
and EN50281-1-1.
In hazardous areas the transmitter has always to be operated via Zener barriers or an approved switch
amplifier! There must be a galvanic separation between the signal lines and the power supply.
We recommend the use of a HygroFlex transmitter. The following types may be used:
HTS 11X, HTS 21X, HTS 31X, HTS 11D, HTS 21D, HTS 31D. barriers!
Further copies of these operating instructions can be obtained from either ROTRONIC AG or one of our representatives.
2. Installation
2.1 Mounting
A
41
150
5 8
2 5 0
150
58
max. 2.0m
28
120
ш 15
270
ш 15
Room type
mounting of the electronic case with 4 screws (not included)
Duct type
Special fittings for through-wall
maounting are available
Type with cable probe
Variant A: mounting like room type.
Variant B: probe fixing with AGRO-screw fitting and flange.
The flange is only necessary when the screw fitting alone cannot
fixed.
Screw in probe
Note:
The screw-in probe must
be fixed together with the
housing in order to avoid a
twisting of the probe cable.
The housing must be fixed only
after screwing the probe in.