8 Specifications

Specifications

473-RP2

473-SH2

Measuring Ranges

Frost/Dew Point
Relative humidity

Temperature (head mount PRT)
Temperature (cable mount PRT)


-

20…70 °C

5…100 %RH
-20...80 °C

-

50…100 °C


-

30…70 °C

2…100 %RH
-

20…80 °C

-

50…100 °C

Accuracy

Frost/Dew point
Temperature


≤ ± 0.1 °C (-20…70 °C), ± 0.2 °C
≤ ± 0.1 °C

Reproducibility

Frost/Dew point
Temperature


≤ ± 0.05 °C
≤ ± 0.05 °C

Standard Features

Temperature probe

Digital I/O

Display
Thermoelectric mirror cooling
Mirror temperature sensor
Gas connections
Transport case
Power cable
Operating instructions
Calibration certificate

RP2: Ø3 x 30 mm PRT, 0.5 m cable
SH2: Ø2.1 x 100 mm PRT, 0.5 and 3 m cables
RS-232
5.7” LCD with color touch screen
2-stage with typically 50 °C depression at 20 °C ambient
Platinum Resistance Thermometer (Pt100)
6 mm or ¼” Swagelok (SH2 only)
Custom fit foam-lined Peli 1550
2.5 m
English
Factory calibration: 5 points FP/DP, 3 points temperature
Upgrade to SCS accredited ISO17025 calibration available

Optional

Int. barometric pressure sensor
Analog outputs

0.1%

or 0.01% accuracy, 700…1200 mbar

User programmable, 2x 4…20 mA or 0…10 V

Additional Information

Power supply
Operating conditions:

Instrument
Measuring heads

Storage conditions

100…120 VAC / 200…240 VAC, 50/60 Hz, 100 Watt (auto switching)

0…40 °C, 90 %RH non-condensing
20…80 °C, 99 %RH non-condensing
-20...50 °C

Weight & Dimensions

Dimensions
Weight
Protection

Instrument

W310 x H155 x D265 mm
5 kg
IP54

In Transport Case

W510 x H220 x D450 mm
12 kg
IP65

We reserve the right to change design or technical data without notice.