beautypg.com

Dwyer TFM-LP User Manual

Page 29

background image

F-20-TF-L, pg. 29 of 30

Symptom

Possible Cause

Method of Correction

Flow indication is not
linear and output too
low at lower flows

Attempting to measure flows
below the specified
minimum for the unit

Liquid is too viscous






Entrapped air or gas


Mechanical sensor
assembly damaged

Use higher flow rates.



Best results are obtained
with liquids that have a
viscosity of 10cS or less.
Contact Dwyer Customer
Service for further
information.

Remove entrapped air (see
section C2).

Unit must be returned for
repair (see Section D3).

Flow indication
“bounces” or
fluctuates

Flow is actually fluctuating


Not using the fittings
supplied by the
manuafacturer

Entrapped air or gas


Turbulence in tubing or
sensor


Fluctuating input power



Mechanical sensor
assembly damaged

Correct flow source to
obtain steady flow.

Use the supplied fittings.



Remove entrapped air (see
section C2).

Straighten the inlet tubing or
lengthen it to at least 6”
(150mm) (see section B3).

Correct the power source or
change to a regulated
supply.

Unit must be returned for
repair (see Section D3).

Flow indication is
fixed at one value

The maximum rated flow for
the unit is being exceeded.

Output circuits may be
damaged

Reduce the flow.


Unit must be returned for
repair (see Section D3).


This manual is related to the following products: