Technical data – West Control Solutions D280-1 User Manual
Page 45
Technical data
Digital 280-1
45
Measuring range
Input resistance
Accuracy
Resolution (
Æ)
0...20 mA
49
W (voltage requirement
£ 2,5 V)
£ 0,1 %
0,75
mA
0...10 Volt
» 110 kW
£ 0,1 %
0,4 mV
-2,5...115 mV*
? 1M[
£ 0,1 %
4 µV
-25...1150 mV*
? 1M[
£ 0,1 %
40 µV
-25...90 mV*
? 1M[
£ 0,1 %
4µV
-500...500 mV*
? 1M[
£ 0,1 %
40 µV
-5...5Volt
» 110 kW
£ 0,1 %
0,4 mV
* high-impedance voltage ranges without break monitoring
Table 3 Current and voltage measuring ranges
Type
Meas.curr.
Measuring range
Accuracy
Resolution (
Æ)
Pt100
0,2 mA
-200...850°C
-328...1562°F
£ 1 K
0,05 K
Pt1000
-200...850°C
-328...1562°F
£ 2 K
Special
*
0...4500 [**
£ 0,1 %
0,005 %
Special
0...450 [**
Pot.
0...160 [**
Pot.
0...450 [**
Pot.
0...1600 [**
* Characteristic KTY 11-6 (-50...150°C) is factory-set.
** inclusive of lead resistance
Table 2 Resistance transducer measuring ranges
thermocouple type
measuring range
accuracy
resolution (
Æ)
L
Fe-CuNi (DIN)
-100...900°C
-148...1652°F
£ 2 K
0,05 K
J
Fe-CuNi
-100...1200°C
-148...2192°F
£ 2 K
0,05 K
K
NiCr-Ni
-100...1350°C
-148...2462°F
£ 2 K
0,1 K
N
Nicrosil/Nisil
-100...1300°C
-148...2372°F
£ 2 K
0,1 K
S
PtRh-Pt 10%
0...1760°C
32...3200°F
£ 2 K
0,1 K
R
PtRh-Pt 13%
0...1760°C
32...3200°F
£ 2 K
0,1 K
T
Cu-CuNi
-200...400°C
-328...752°F
£ 2 K
0,025 K
C
W5%Re-W26%Re
0...2315°C
32...4199°F
£ 2 K
0,2 K
D
W3%Re-W25%Re
0...2315°C
32...4199°F
£ 2 K
0,2 K
E
NiCr-CuNi
-100...1000°C
-148...1832°F
£ 2 K
0,05 K
B
(1)
PtRh-Pt6%
0(100)...1820°C
32(212)...3308°F
£ 3 K
0,15 K
special
-25....75 mV
£ 0,1 %
0,005 %
(1)
Values for type B are valid from 100°C.
Tabelle 1 thermocouple measuring range