16 ADDENDUM 4: METER FACTORY DEFAULT

SETTINGS

Type Parameter

Default

Remarks

P1.0

Memory clear

No

Retains current memory

P2.1

Viewing previous

––

No calibration data for 1

st

range

P2.2

calibration data

––

No calibration data for 2

nd

range

P2.3

––

No calibration data for 3

rd

range

P2.4

––

No calibration data for 4

th

range

P2.5

––

No calibration data for 5

th

range

P3.1

Viewing probe data

1.0

No offset for effective cell constant (1

st

range)

P3.2

1.0

No offset for effective cell constant (2

nd

range)

P3.3

1.0

No offset for effective cell constant (3

rd

range)

P3.4

1.0

No offset for effective cell constant (4

th

range)

P3.5

1.0

No offset for effective cell constant (5

th

range)

P4.1

Ready indicator / auto
endpoint

READY/ON Ready

indicator on; auto endpoint off

P4.2 Select

°C/°F

°C

Degrees

Celsius

P4.3

ATC on or off

ATC on

––

P4.4

TDS factor

0.50

Adjustable from 0.40 to 1.00 (available in TDS
400 and CON 410 meters only
)

P5.1

Temperature coefficient

2.1 % per °
C

Adjustable from 0.0 to 10%

P5.2

Normalisation temperature

25° C

Adjustable from 15 to 30° C

P6.1

Cell constant

1.0

Select from k=1.0, 0.1 or 10

P8.0

Factory default

No

Retains your current settings


NOTES:
The P7.0 clock function retains the year and time that you set, even after reverting to factory
default settings.