TROUBLE SHOOTING GUIDE

Problem Cause

Solution

No display when
turned ON

a) Batteries not in place or

battery housing is not
attached to meter correctly

b) Batteries not in correct

polarity (+ and – position).

c) Weak

batteries

a) Check that batteries are in

place and making good
contact.

b) Re-insert batteries with

correct polarity.

c) Replace batteries.

Unstable readings

a) Air bubbles in probe.

b) Dirty probe.

c) Probe not deep enough in

sample.

d) External noise pickup or

induction caused by nearby
electric motor.

e) Broken probe.

f) probe guard not in place

a) Tap probe to remove

bubbles.

b) Clean the probe and re-

calibrate.

c)

Make sure sample entirely
covers the probe sensors.

d) Move or switch off

interfering motor.

e) Replace probe.

f)

Attach probe guard

Slow response

a) Dirty / Oily probe.

a) Clean probe. See “Probe

Care & Maintenance”, page
40.