beautypg.com

Thermo Fisher Scientific CyberScan CON 400 (For units manufactured from March 2010 onwards) User Manual

Page 53

background image

Instruction Manual

CON 400

- 47 -

15 ADDENDUM 3: CALCULATING TEMPERATURE

COEFFICIENTS

To determine the temperature coefficient of your sample solution use this formula:

Where:

tc = Temperature coefficient

25 = 25 °C

C

T1

= Conductivity at Temp 1

C

T2

= Conductivity at Temp 2

T

1

= Temp 1

T

2

= Temp 2


NOTE:
A controlled temperature water bath is ideal for this procedure.

1. Immerse the probe into a sample of your solution and adjust the temperature coefficient

to 0% (that is, no compensation) by performing the following:
A. From measurement mode, press the SETUP key to enter Setup mode.
B. Press

the

MI/S or MR/T key until the lower display reads P5.0.

C. Press

the

ENTER key twice. The lower display reads tCO and the upper display

shows the temperature coefficient value.

D. Press

the

MR/T key until the upper display shows 0.0.

E. Press

ENTER key to confirm the value.

F. Press

CA/MEAS key twice to return to measurement mode.

2.

Wait for 5 minutes. Note T

1

and C

T1

(conductivity at T

1

).

3.

Condition the sample solution and probe to a temperature (T

2

) that is about 5 °C to 10 °C

different from T

1

, and note the conductivity reading C

T2

.


NOTE:
Record your results for future reference. Ideally T

1

and T

2

should bracket your

measurement temperature, and should not different by more than 5 °C.


4. Calculate the temperature coefficient of your solution according to the formula shown

above.

5.

Enter the temperature coefficient you calculated into the meter.
A. From measurement mode, press the SETUP key to enter Setup mode.
B. Press

the

MR/T key until the lower display reads P5.0.

C. Press

the

ENTER key twice. The lower display reads tCO and the upper display

shows the temperature coefficient value (should be 0, as per step 1 above).

D. Press

the

MI/S key until the upper display shows your calculated temperature

coefficient.

E. Press

ENTER key to confirm the value.

F. Press

CAL/MEAS key twice to return to measurement mode.


The calculated temperature coefficient will not be applied to all the meter readings.