Instruction Manual

CON 400

- 16 -

5.2 Manual

Temperature

Compensation

IMPORTANT: For manual compensation, you must deactivate the automatic temperature
compensation feature.

5.2.1

Selecting Manual Temperature
Compensation

To select between Automatic Temperature Compensation
(ATC) and Manual Temperature Compensation in the
SETUP program P4.3. Meter default is ATC on.

From measurement mode
1) Press

SETUP key to enter Set Up mode.

2) Press

the

MI/S or MR/T keys to scroll through

subgroups until you view parameter P4.0.

3) Press

the

ENTER key three times to select parameter

4.3. The upper display shows “ATC” and the lower
display shows “P4.3”.

4) Press

the

ENTER key again. The upper display shows

“ATC” and the lower display shows “YES” or “NO”.

5) Press

the

MI/S or MR/T to select between Automatic

or Manual Temperature Compensation feature.
• YES = Automatic Temperature Compensation

activated.

• NO = Manual Temperature Compensation

activated.

6) Press

the

ENTER key to confirm selection and to

return to the subgroup menu. Press the CAL/MEAS
key to return to measurement mode.


NOTE:
Manual Temperature Compensation should only be
selected if the ATC fails. For non-compensated
measurements, change the temperature coefficient to
0.0% (see section 7.6.1)

P4.0 Configuration Setup

Selecting Automatic or Manual

Temperature Compensation