Thermo Fisher Scientific CyberScan CON 400 (For units manufactured from March 2010 onwards) User Manual
Page 22

Instruction Manual
CON 400
- 16 -
5.2 Manual
Temperature
Compensation
IMPORTANT: For manual compensation, you must deactivate the automatic temperature
compensation feature.
5.2.1
Selecting Manual Temperature
Compensation
To select between Automatic Temperature Compensation
(ATC) and Manual Temperature Compensation in the
SETUP program P4.3. Meter default is ATC on.
From measurement mode
1) Press
SETUP key to enter Set Up mode.
2) Press
the
MI/S or MR/T keys to scroll through
subgroups until you view parameter P4.0.
3) Press
the
ENTER key three times to select parameter
4.3. The upper display shows “ATC” and the lower
display shows “P4.3”.
4) Press
the
ENTER key again. The upper display shows
“ATC” and the lower display shows “YES” or “NO”.
5) Press
the
MI/S or MR/T to select between Automatic
or Manual Temperature Compensation feature.
• YES = Automatic Temperature Compensation
activated.
• NO = Manual Temperature Compensation
activated.
6) Press
the
ENTER key to confirm selection and to
return to the subgroup menu. Press the CAL/MEAS
key to return to measurement mode.
NOTE:
Manual Temperature Compensation should only be
selected if the ATC fails. For non-compensated
measurements, change the temperature coefficient to
0.0% (see section 7.6.1)
P4.0 Configuration Setup
Selecting Automatic or Manual
Temperature Compensation