7.5 P4.0:

Unit

configuration

7.5.1

P4.1: READY indicator and auto
endpoint function

Program P4.1 lets you select “READY indicator on” to
indicate when your measurement is stable, or select
“READY indicator off” for faster meter response.

Program P4.1 also lets you switch the Auto endpoint
function on or off. Select auto endpoint on to “hold” your
measurement when it is stable for more than 5 seconds.
The display automatically freezes, and HOLD indicator
appears on the left side of the display. Press the HOLD
key to release the display and access other functions.
Select auto endpoint off to deactivate this feature.

From measurement mode
1. Press

SETUP key to enter Set Up mode.

2. Press

the

MI/S or MR/T keys to scroll through

subgroups until you view parameter P4.0.

3. Press

ENTER key to select parameter 4.1.

4. Press

the

MI/S or MR/T keys to select the

configuration you require.
• OFF switches the READY indicator off; ON

switches the READY indicator on. ON and HOLD
together switches the auto endpoint feature on.

5. Press

ENTER key to confirm selection and to

proceed to step 4 of P4.2. You can also press the
CAL/MEAS key to return back to measurement
mode.


NOTE: Meter default is set for Ready Indicator on, and auto
endpoint function off.

Figure 26 - Configuring the

READY feature