Thermo Fisher Scientific CyberScan CON 400 (For units manufactured from March 2010 onwards) User Manual
Page 34

Instruction Manual
CON 400
- 28 -
7.3
P2.0: Viewing calibration data
This mode lets you recall previous calibration data, which
helps you know when to re-calibrate your meter. This is
a “view only” mode. See Figure 24.
From measurement mode:
1. Press
the
SETUP key to enter Set up mode.
2. Press
the
MI/S or MR/T keys to scroll through
subgroups until you view parameter P2.0.
3. Press
ENTER key repeatedly to view previous
calibration data. The meter will first display the
calibration point, and then display the date and time
of calibration to help you comply with the Good
Laboratory Practice (GLP) standards.
4.
When you have scrolled through all calibration data,
you will automatically return to the subgroup menu.
Press CAL/MEAS key to return to measurement
mode.
NOTES:
If there is no previous calibration data at a particular point,
the primary display will show “ ----“.
If you entered Set Up mode from Conductivity
measurement mode, calibration data will be in
μS or mS.
If you entered Set Up mode from TDS measurement
mode, calibration data will be in ppm or ppt.
Figure 24 - View calibration data
and its relevant time and date