beautypg.com

Thermo Fisher Scientific CyberScan CON 400 (For units manufactured from March 2010 onwards) User Manual

Page 19

background image

Instruction Manual

CON 400

- 13 -

4.5.2

Calibration with Conductivity Standard and TDS factor


The concentration of salts dissolved in solution increases the conductivity of that solution. This
relationship varies from salt to salt and is roughly linear over a given range for a given salt.
The TDS conversion factor is the number used by the meter to convert from conductivity to
TDS.

Instead of calibrating for TDS directly (described in Section 4.5.1), you can calibrate the CON
400 meter by:
1.

calibrating to conductivity standards (as described on pages 10 to 11) and then

2.

entering the appropriate TDS conversion factor into the meter.


To determine the conductivity to TDS conversion factor for your solution:

• Addendum 2 explains the formula to calculate the conductivity-to-TDS

conversion factor.


Enter the TDS conversion factor into your meter as described under Section 7.5.4, in P4.4
Setting the TDS Factor
on page 33.