Instruction Manual

CON 400

- 36 -

7.7

P6.0 Selecting the cell constant

Your meter lets you select a cell constant of K = 1.0, 10, or 0.1.
Use a cell of K = 1.0 for midrange measurements
Use a cell of K = 10 for high range measurements (above 20 mS or 10 ppt).
Use a cell of K = 0.1 for low range measurements (below 20

μS or 10 ppm).

The cell included with your meter has a cell constant of K = 1.0.

From measurement mode
1. Press

SETUP key to enter Set Up mode.

2. Press

the

MI/S or MR/T keys to scroll through

subgroups until you view parameter P6.0.

3. Press

the

ENTER key to select parameter 6.1.

4. Press

the

MI/S or MR/T keys to select the cell

constant between K = 1.0, 0.1, or 10.

5. Press

the

ENTER key to confirm selection and to

return to the subgroup menu. Press the CAL/MEAS
key to return to measurement mode.





NOTES
When using a cell of K = 0.1, the lowest measuring range
will be: 0 to 1.999

μS (0 to 0.999 ppm). The 0 to 199.9 ms

(10 to 200 ppt) range will not be accessible.
When using a cell of K = 10, the highest measuring range
will be: 0 to 1999 mS (100 to 2000 ppt). The 0 to 19.99

μS

(0.00 to 9.99 ppm) range will not be accessible.

Figure 32 - Selecting cell constant