13 ADDENDUM 1: CALIBRATION TIPS

You only need one calibration for measurement throughout the entire range of

the meter. If a range was not calibrated, the meter automatically detects the

closest range calibrated and uses that calibration information. However, only

the ranges that were calibrated have maximum accuracy.

If you are measuring in ranges near to or greater than 20 mS (10 ppt), or near

to or lower than 100 µS (50 ppm), calibrate the meter at least once a week to

get specified ±1% F.S. accuracy.
If you are measuring in the mid-ranges and you washed the probe in deionized

water and stored it dry, calibrate the meter at least once a month.

Wet the probe for 10 minutes before calibrating or taking readings to saturate

the probe surface and minimize drift. If you make measurements at extreme

temperatures, calibrate the meter at least once a week.

You should only use the conductivity / TDS probe specified for these meters.

These probes have a built-in temperature sensor. If you use a different probe

without a temperature sensor, you must measure the solution temperature

separately and manually enter the solution temperature (see manual

temperature compensation section 5.2)