Instruction Manual CON 6/TDS 6

43

Some users prefer the use of resistivity units to describe their water, particularly

where high purity water is involved. The unit most often used to describe

resistivity is megohm-cm, which are simply the reciprocal of conductivity

(µS/cm). The chart below shows the relationship between these units.

Conductivity, µS/cm

Resistivity, megohm-cm

0.056

18

0.1

10

1.0

1.0

2.5

0.4

10.0

0.1