Instruction Manual CON 6/TDS 6

29

6.2

Select Cell Constant

This meter lets you select a cell constant of K = 1.0, 10,

or 0.1.

Use a cell of K = 1.0 for midrange

measurements

Use a cell of K = 10 for high range

measurements (above 20 mS or 10 ppt).

Use a cell of K = 0.1 for low range

measurements (below 20 µS or 10 ppm).

The cell included with your meter has a cell constant of

K = 1.0.
1)

Enter the advanced setup as described in

Section 6.1.

2)

Press ▲ or ▼ key until [CELC] appears on the

LCD. Press ENTER key.

3)

Press ▲ or ▼ key to select either “1.0”, “0.1” or

“10.0”. Ensure the cell constant selected

correspond with the conductivity electrode you are

using.

4)

Press ENTER key to select. The meter will take

you back to the menu, [CELC].

5)

Press ▲ or ▼ key to move to the next menu or

press CAL to exit to measurement mode.