Thermo Fisher Scientific EcoScan CON 6 & TDS 6 User Manual

Page 15

background image

Instruction Manual CON 6/TDS 6

11

Press ON/OFF key to power up your meter.
1)

The first screen shows [Con 6] (or [tdS 6]) which is the meter‟s name.

2)

Second screen shows [C 1.0] which is the conductivity cell constant, k.

You can select different cell constant of 0.1, 1.0 or 10.0. Refer to Section

on Advance Setup. Default value is k=1.0.

3)

Third screen shows [t 25.0 °C] which is the Normalization Temperature.

You can set Normalization Temperature at either 25 °C or 20 °C. Refer to

Section on Advance Setup. Default value is 25 °C.

4)

Fourth screen shows [t 2.1%] which is the Temperature Coefficient. You

can customise the meter with different Temperature Coefficient value

from 0.0 to 3.0 %/°C from the Advance Setup mode. Default value is 2.1

%/°C.

5)

All LCD segments will light up for 2 seconds, and change into

measurement mode.

6)

You are now ready for conductivity measurement.