Instruction Manual CON 6/TDS 6

32

6.6

Normalization Temperature

You can set the meter to normalize its conductivity measurements to a standard

temperature of either 25 °C or 20 °C.
The default value is 25 °C.
1)

Enter the advanced setup as described in

Section 6.1.

2)

Press ▲ or ▼ key until [t.nr °C] appears on the

LCD. Press ENTER key.

3)

Press ▲ or ▼ key to select either [25.0 °C] or

[20.0 °C].

4)

Press ENTER key to select. The meter will take

you back to the menu, [t.nr °C].

5)

Press ▲ or ▼ key to move to the next menu or

press CAL to exit to measurement mode.

6.7

Single-Point Calibration

Single-point calibration refers to calibrating one conductivity value and uses it

for the entire 5 conductivity ranges.
By selecting [no] to single-point calibration, you can perform calibration for each

conductivity range.
1)

Enter the advanced setup as described in

Section 6.1.

2)

Press ▲ or ▼ key until [S.P.CA] appears on the

LCD. Press ENTER key.

3)

Press ▲ or ▼ key to select either [Yes] or no].

4)

Press ENTER key to select. The meter will take

you back to the menu, [S.P.CA].

5)

Press ▲ or ▼ key to move to the next menu or

press CAL to exit to measurement mode.