3 automatic calibration (for con 6), 4 setting the tds factor (for tds 6) – Thermo Fisher Scientific EcoScan CON 6 & TDS 6 User Manual

Page 34

background image

Instruction Manual CON 6/TDS 6

30

6.3

Automatic Calibration (for CON 6)

The automatic calibration allows you to quickly calibrate the meter to any of the

four widely used conductivity calibration standards. For a list of calibration

standards refer to Table 1 in Section 4.3.
In the manual calibration mode, you can use you own customized conductivity

calibration standard to calibrate this meter.
1)

Enter the advanced setup as described in

Section 6.1.

2)

Press ▲ or ▼ key until [ACAL] appears on the

LCD. Press ENTER key.

3)

Press ▲ or ▼ key to select either [Yes] or [no].

4)

Press [ENTER] key to select. The meter will take

you back to the menu, [ACAL].

5)

Press ▲ or ▼ key to move to the next menu or

press CAL to exit to measurement mode.

6.4

Setting the TDS Factor (for TDS 6)

The concentration of salts dissolved in solution

increases the conductivity of that solution. This

relationship varies from salt to salt and is roughly

linear over a given range for a given salt. The TDS

conversion factor is the number used by the meter to

convert from conductivity to TDS.
To calculate the TDS conversion factor refers to

Addendum 2 – Calculating TDS Conversion Factor.
You can also look up at various Chemical reference

books for TDS factor for various types of salt.
You can set the TDS conversion factor between 0.4

and 1.0; meter default is 0.5.
1)

Enter the advanced setup as described in Section 6.1.

HOLD

ENTER

HOLD

ENTER