beautypg.com

2 preparing the meter for calibration – Thermo Fisher Scientific EcoScan CON 6 & TDS 6 User Manual

Page 18

background image

Instruction Manual CON 6/TDS 6

14

4.2

Preparing the Meter for Calibration

Before starting calibration, make sure you are in the correct measurement

mode.
For best results, select a standard value close to the sample value you are

measuring. Alternatively use a calibration solution value that is approximately

2/3 the full-scale value of the measurement range you plan to use. For

example, in the 0 to 2000 µS conductivity range, use a 1413 µS solution for

calibration.
Calibrate to all measurement ranges to ensure the highest accuracy throughout

all measurement range. Note that CON 6/ TDS 6 will not accept calibration

values of less than 2µS/cm during the multi-point calibration process. All new

calibration values will automatically override existing data.
If you are measuring in solutions with Conductivity lower than 100 µS/cm or

TDS lower than 50 ppm, calibrate the meter at least once a week to get good

accuracy. If you are measuring in the mid ranges and you wash the probe in

deionized water and store it dry, calibrate the meter once a month. If you take

measurements at extreme temperatures, calibrate at least once a week.
Ensure that you use new Conductivity standard solutions or sachets during

calibration. Do not reuse standard solutions as it may be contaminated and

affect the calibration and accuracy of measurements. Use fresh calibration

solution each time you calibrate your meter. Store solutions in a dry and cool

environment if possible.
Always rinse the probe with either deionized water or rinse solution before and

after each calibration/sample measurement to avoid cross-contamination. For

details please refer to Section 7 - Probe Care and Maintenance.

NOTE: These meters are factory set to a temperature

coefficient of 2.1% per °C. For most applications

this will provide good results.

To set the temperature coefficient to different value,

see Section 6.5 – Temperature Coefficient.

Also, see Addendum 3 - Calculating Temperature

Coefficient to determine the appropriate

temperature coefficient for your solution.