Instruction Manual CON 6/TDS 6

19

4.6

Temperature Calibration

The Conductivity electrode (Refer to Accessories section

for order number) has a built-in temperature sensor for

ATC. The temperature sensor is factory calibrated to the

meter. Calibrate your sensor only if you suspect

temperature errors that may have occurred over a long

period of time or if you have a replacement probe.
1)

Make sure that the phono jack (for temperature

measurement) is properly connected to the meter.

See Figure in Section 3.5.

2)

Switch on the meter and if necessary, press MODE

key to select temperature measurement mode. See

Section 3.7.

3)

Press CAL key to start temperature calibration

process.

4)

Dip the probe into a solution with known

temperature (for example, a temperature bath).

Allow time for the temperature to stabilize.

5)

Wait for the value to stabilize and press ▲ or ▼ key

and adjust the value to the solution temperature.

6)

Press the ENTER key. The [CO] indicator will

appear for 1.5 seconds, and the reading will stop

flashing. The temperature calibration is successfully

performed. The meter returns to measurement

mode.

NOTE: To exit calibration without confirmation, press CAL key again to go

back to measurement mode.

NOTE: You can offset the temperature reading up to ±5 °C from

the original (default) reading.