6 temperature calibration – Thermo Fisher Scientific EcoScan CON 6 & TDS 6 User Manual
Page 23

Instruction Manual CON 6/TDS 6
19
4.6
Temperature Calibration
The Conductivity electrode (Refer to Accessories section
for order number) has a built-in temperature sensor for
ATC. The temperature sensor is factory calibrated to the
meter. Calibrate your sensor only if you suspect
temperature errors that may have occurred over a long
period of time or if you have a replacement probe.
1)
Make sure that the phono jack (for temperature
measurement) is properly connected to the meter.
See Figure in Section 3.5.
2)
Switch on the meter and if necessary, press MODE
key to select temperature measurement mode. See
Section 3.7.
3)
Press CAL key to start temperature calibration
process.
4)
Dip the probe into a solution with known
temperature (for example, a temperature bath).
Allow time for the temperature to stabilize.
5)
Wait for the value to stabilize and press ▲ or ▼ key
and adjust the value to the solution temperature.
6)
Press the ENTER key. The [CO] indicator will
appear for 1.5 seconds, and the reading will stop
flashing. The temperature calibration is successfully
performed. The meter returns to measurement
mode.
NOTE: To exit calibration without confirmation, press CAL key again to go
back to measurement mode.
NOTE: You can offset the temperature reading up to ±5 °C from
the original (default) reading.