Instruction Manual CON 6/TDS 6

18

4.5

Manual Calibration (Conductivity or TDS)

In Manual Calibration mode, you can use customized conductivity calibration

standards (specific to your own application) and calibrate the meter. The

following example shows calibration sequence to 12.00 mS conductivity

calibration standard.
Procedure is similar for CON 6 and TDS 6 meters.
1)

If necessary, press MODE key to select conductivity

mode.

2)

Rinse the probe thoroughly with de-ionized water or

a rinse solution, then rinse with a small amount of

calibration standard.

3)

Dip the probe into the calibration standard. Immerse

the probe tip beyond the upper steel band (see

Figure in Section 3.4). Stir the probe gently to

create a homogeneous sample. Allow time for the

reading to stabilize.

4)

Press CAL key to enter conductivity calibration

mode. The [CA] indicator will appear for 1.5

seconds, and a value will appear flashing.

NOTE: To exit calibration without confirmation, press

CAL key again to go back to measurement
mode.

5)

Wait for the value to stabilize and press ▲ or ▼ key

and adjust the value to the calibration standard

used.

6)

Press the ENTER key. The [CO] indicator will appear for 1.5 seconds,

and the calibration is successfully performed. The meter returns to

measurement mode.

7)

To perform the next point calibration in the multi-point calibration for next

range, repeat step 1-6 again until all points have been calibrated if

necessary.