SHIMPO DT-329 User Manual
Page 28

25
4.6 Connector of External input/output
4.6.1 Connector Pin Assign
External Input
Pin No.
Signal Name
Contents
1
+12V
+12V power supply output
2
G12
Ground of power supply output
3
TRG_IN
External pulse input
4
IN_COM
Common of External pulse input
5
NC
6
NC
7
FG
Frame ground
DT-329 side: SRCN2A16-7S [JAE]
Accessory: SRCN6A16-7P [JAE]
External Output
Pin No.
Signal Name
Contents
1
OUT
External pulse output
2
OUT_COM
Common of External pulse output
3
FG
Frame ground
DT-329 side: SRCN2A13-3S [JAE]
Accessory: SRCN6A13-3P [JAE]
4.6.2 External Pulse Input
The unit can be operated and controlled by an external signal, from either a sensor or a machine signal like
an output from a printing press, re-winder or others. This will allow the stroboscope to flash in sync with the
target object’s rotational speed.
The input signal will have the following characteristics.
Input frequency: 60 - 10,000 FPM (1 – 160 Hz)
Input pulse width: over 50us
Delay angle: 0 - 359°
(
every 1° )
Delay time: 0 - 999ms (every 1ms)
Please connect with open collector output (below 1V (ON voltage)) or no-voltage contact output (approx.
11mA current)
Connector of
External
input
3pin
4pin
+12V
Inside DT-329
Photocoupler
1k
Ω
[Input circuit]