12. Specifications
12.1 General Specifications
Warm up time
Twice the time since last warmed up, to a maximum of 30 minutes.
Settling time
Less than 5 seconds for all functions and ranges except as noted.
Standard interfaces
RS-232
IEEE-488 (GPIB)
Temperature performance
Operating
0 °C to 50 °C
Calibration (tcal)
18 °C to 28 °C
Storage
-20 °C to 70 °C
Electromagnetic compatibility
CE: Conforms to EN61326
Temperature coefficient
Temperature coefficient for temperatures outside tcal ±5 °C is 10% of
the 90 day specification (or 1 year if applicable) per °C
Relative humidity
Operating
<80% to 30 °C
<70% to 40 °C
<40% to 50 °C
Storage
<95 % noncondensing
Altitude
Operating 3,050 m (10,000 ft) maximum
Nonoperating 12,200 m (40,000 ft) maximum
Safety
EN 61010 Second, ANSI/ISA-S82.01-1994,
CAN/CSA-C22.2 No. 1010.1-92, NRTL
Analog low isolation
20 V
Line power
Line Voltage (selectable)
100V/120V or 220V/240V
Line Frequency
47 to 63 Hz
Line Voltage Variation
±10 % about setting
Power consumption
15 VA maximum
Dimensions
Height
13.3 cm (5.25 in) plus 2.9 cm (1.15 in) for extended feet
Width
standard rack width (19 inch)
Depth
30.0 cm (11.81 in) overall
Weight (without options)
4 kg (9 lb)
79