Martel Electronics 3001 User Manual
Page 56
Measurement Commands
Command
Description
ISO_MEAS
Sets the isolated input measurement type.
PRES?
Queries the attached pressure module for its manufacturer and serial num-
ber.
PRES_MEAS
Changes the operating mode of the primary display to pressure measure-
ment.
RTD_MEAS
Changes the operating mode to RTD measurement.
TC_MEAS
Changes the operating mode to thermocouple measurement.
VAL?
Returns the last values for the isolated and primary measurements, in that
order.
ZERO_MEAS
Zeros the pressure module, TC mV, or RTD ohms.
ZERO_MEAS?
Returns the zero offset for the pressure module, TC mV, or RTD ohms.
RS-232 Operating Mode Commands
Command
Description
LOCAL
Puts the 3001 into the local state and disables lockout.
LOCKOUT
Puts the 3001 into the lockout state. This command duplicates the IEEE-488
LLO (Local Lockout) message.
REMOTE
Puts the 3001 into the remote state. This command duplicates the IEEE-488
REN (Remote Enable) message.
Status Commands
Command
Description
FAULT?
Returns the most recent error code in the 3001 error queue, and then
removes that error code from the queue.
10.3 Error Code Listing
Error Number Message Class
Description
1
DDE
Error queue overflow.
100
EXE
OPER or STBY was received when the 3001 is in measure
mode or thermocouple source mode.
101
CME
A non-numeric entry was received in a field that should con-
tain a numeric entry.
102
EXE
The numeric field exceeds 10 characters.
103
CME
Invalid units name or prefix.
104
EXE
An attempt to enter RTD source mode was made when SPRT
is selected, or to select SPRT when RTD source mode is
selected.
105
EXE
Entry is above upper limit for the selected output range.
106
EXE
Entry is below lower limit for the selected output range.
52