Martel Electronics 3001 User Manual

Page 57

background image

108

CME

A required command parameter was missing.

109

CME

An invalid TC_MEAS or RTD_MEAS unit parameter (not CEL
or FAR) was received, or an invalid PRES_UNIT or
ISO_PRES_UNIT parameter was received.

110

CME

An invalid RANGELCK parameter was received.

111

EXE

RANGELCK ON was received when the 3001 is not in Volts
mode.

112

CME

An invalid RTD_TYPE parameter was received.

113

CME

An invalid TC_REF parameter was received.

114

CME

An invalid TSENS_TYPE parameter was received.

116

EXE

No pressure module was present when a pressure function
was requested, or the total zero offset from calibration is more
than 6%. This error can also occur if the total zero offset from
calibration is out of limits when zeroing the thermocouple mil-
livolts type (maximum offset ±1 mV), or the RTD ohms type
(maximum ±0.1 ohm on high range, or ±0.01 ohm on the low
range).

117

CME

An unrecognizable command was received.

118

CME

An invalid parameter was received.

120

EXE

The serial input buffer overflowed

121

EXE

The command string buffer overflowed.

122

QYE

The serial output buffer overflowed.

123

DDE

The output overloaded. See display error message OVER
LOAD in section 2.7.

124

DDE

The 3001 is out of tolerance. This error is set after a failed ini-
tialization or a failed *TST? command.

125

DDE

The 3001 ADC has failed. This error is set after a failed initial-
ization or a failed *TST? command.