Martel Electronics 3001 User Manual
Page 57
108
CME
A required command parameter was missing.
109
CME
An invalid TC_MEAS or RTD_MEAS unit parameter (not CEL
or FAR) was received, or an invalid PRES_UNIT or
ISO_PRES_UNIT parameter was received.
110
CME
An invalid RANGELCK parameter was received.
111
EXE
RANGELCK ON was received when the 3001 is not in Volts
mode.
112
CME
An invalid RTD_TYPE parameter was received.
113
CME
An invalid TC_REF parameter was received.
114
CME
An invalid TSENS_TYPE parameter was received.
116
EXE
No pressure module was present when a pressure function
was requested, or the total zero offset from calibration is more
than 6%. This error can also occur if the total zero offset from
calibration is out of limits when zeroing the thermocouple mil-
livolts type (maximum offset ±1 mV), or the RTD ohms type
(maximum ±0.1 ohm on high range, or ±0.01 ohm on the low
range).
117
CME
An unrecognizable command was received.
118
CME
An invalid parameter was received.
120
EXE
The serial input buffer overflowed
121
EXE
The command string buffer overflowed.
122
QYE
The serial output buffer overflowed.
123
DDE
The output overloaded. See display error message OVER
LOAD in section 2.7.
124
DDE
The 3001 is out of tolerance. This error is set after a failed ini-
tialization or a failed *TST? command.
125
DDE
The 3001 ADC has failed. This error is set after a failed initial-
ization or a failed *TST? command.
53