For these commands, the detailed command descriptions in section 10 show a
check mark
beside RS-232, but no check mark
beside IEEE-488.
Table 6 - Commands for RS-232 Only
IEEE-488 Message
RS-232 Equivalent
GTL
LOCAL command
GTR
REMOTE command
LLO
LOCKOUT command
i)
Commands for IEEE-488
These are all of the commands except for those used for RS-232 only, as
described above. All commands are transferred over the IEEE-488 as data,
except for the commands LOCAL, REMOTE, and LOCKOUT, which are
implemented per the IEEE Standards as uniline messages.
The detailed command descriptions in section 10 show a check mark T beside
IEEE-488 for these commands.
9.6.2 Command Syntax
The following syntax rules apply to all of the remote commands. Information about the
syntax of response messages is also given.
a)
Parameter Syntax Rules
Table 7 lists the units accepted in command parameters and used in
responses. All commands and units may be entered in upper or lower case.
Table 7 - Units Accepted in Parameters and Used in Responses
Units Meaning
uV
Volts in units of microvolts1
mV
Volts in units of millivolts1
V
Volts in units of volts
kV
Volts in units of kilovolts1
uA
Current in units of microamperes1
mA
Current in units of milliamps1
A
Current in units of amps
Ohm
Resistance in units of ohms
kOhm
Resistance in units of kilohms1
MOhm
Resistance in units of megohms1
cel
Temperature in degrees Celsius
far
Temperature in degrees Fahrenheit
psi
Pressure in pounds per square inch
mmHg
Pressure in millimeters of mercury at 0 °C
inHg
Pressure in inches of mercury at 0 °C
X
43