beautypg.com

Martel Electronics 3001 User Manual

Page 60

background image

FUNC?

IEEE-488

RS-232

Sequential

Overlapped

This command returns the present output, measurement, or calibration function for the
primary and isolated displays.

Parameter:

Response:

,
where is one of the following:

DC10V

measure DC voltage, 10V range

DC100V

measure DC voltage, 100V range

DCI

measure DC current

PRESSURE measure

pressure

and where is one of the following:

DCV

source DC voltage

DCI

source DC current

RTD_OUT

source RTD/Ohms

RTD_IN

measure RTD/Ohms

TC_OUT

source thermocouple

TC_IN

measure thermocouple

PRESSURE measure

pressure

Example:

FUNC?

DC10V,PRESSURE

This example indicates that the isolated display is selected to the 10V
range and the primary display is selected to pressure.

HART?

IEEE-488

RS-232

Sequential

Overlapped

This command returns the isolated DC current input HART resistor status.

Parameter:

Response:


where is ON or OFF

Example:

HART?

OFF

This example indicates that the isolated DC current input HART resistor
is turned off.

X

X

X

X

X

X

56