Martel Electronics 3001 User Manual
Page 63
![background image](/manuals/369770/63/background.png)
ISO_PRES_UNIT?
IEEE-488
RS-232
Sequential
Overlapped
This command returns the isolated pressure unit.
Parameter:
Response:
where
PSI
pounds per square inch
INH2O4C
inches of water at 4 °C
INH2O20C
inches of water at 20 °C
INH2O60F
inches of water at 60 °F
CMH2O4C
centimeters of water at 4 °C
CMH2O20C
centimeters of water at 20 °C
MMH2O4C
millimeters of water at 4 °C
MMH2O20C
millimeters of water at 20 °C
BAR
bars
MBAR
millibars
KPA
kilopascals
MPA
megapascals
INHG
inches of mercury at 0 °C
MMHG
millimeters of mercury at 0 °C
KG/CM2
kilograms per square centimeter
Example:
ISO_PRES_UNIT?
BAR
This example indicates that the isolated pressure unit is bars.
LOCAL
IEEE-488
RS-232
Sequential
Overlapped
This command puts the 3001 into the local state, clearing the remote state (see the
REMOTE command) and the front panel lockout state (see the LOCKOUT command). It
duplicates setting the IEEE-488 REN line to false.
Parameter:
Response:
Example:
LOCAL
This example puts the instrument into the local state, clearing the
remote state and front panel lockout state, if enabled.
X
X
X
X
X
59