Martel Electronics 3001 User Manual
Page 62
ISO_MEAS
IEEE-488
RS-232
Sequential
Overlapped
This command sets the isolated measurement type.
Parameter:
where
DC10V
measure DC voltage, 10V range
DC100V
measure DC voltage, 100V range
DCI
measure DC current
PRESSURE measure
pressure
Response:
Example:
ISO_MEAS DCI
This example sets the isolated measurement to DC current.
ISO_PRES_UNIT
IEEE-488
RS-232
Sequential
Overlapped
This command sets the isolated pressure unit.
Parameter:
where
PSI
pounds per square inch
INH2O4C
inches of water at 4 °C
INH2O20C
inches of water at 20 °C
INH2O60F
inches of water at 60 °F
CMH2O4C
centimeters of water at 4 °C
CMH2O20C
centimeters of water at 20 °C
MMH2O4C
millimeters of water at 4 °C
MMH2O20C
millimeters of water at 20 °C
BAR
bars
MBAR
millibars
KPA
kilopascals
MPA
megapascals
INHG
inches of mercury at 0 °C
MMHG
millimeters of mercury at 0 °C
KG/CM2
kilograms per square centimeter
Response:
Example:
ISO_PRES_UNIT BAR
This example sets the isolated pressure unit to bars.