Martel Electronics 3001 User Manual
Page 76
TC_REF?
IEEE-488
RS-232
Sequential
Overlapped
This command returns the source of the temperature being used for cold junction
compensation of thermocouple source and measurement.
Parameter:
Response:
where
INT
internal temperature sensor in use
EXT
external reference value in use
Example:
TC_REF?
INT
This example indicates that the internal sensor is in use.
TC_TYPE
IEEE-488
RS-232
Sequential
Overlapped
This command sets the Thermocouple (TC) sensor type being used for TC temperature
source and measurement. Normally, before using the TC_TYPE command to select the
TC type, use the TSENS_TYPE command to select TC mode, and afterwards set the
output temperature using the OUT command, if applicable. A change in temperature
sensors sets the output to 0 °C.
Parameters:
where
B
B-type thermocouple
C
C-type thermocouple
E
E-type thermocouple
J
J-type thermocouple
K
K-type thermocouple
L
L-type thermocouple
N
N-type thermocouple
R
R-type thermocouple
S
S-type thermocouple
T
T-type thermocouple
U
U-type thermocouple
XK
XK-type thermocouple
BP
BP-type thermocouple
Z
1 mV/°C
Response:
Example:
TC_TYPE J
This example sets the thermocouple type to J-type.
X
X
X
X
X
X
72