beautypg.com

Martel Electronics 3001 User Manual

Page 76

background image

TC_REF?

IEEE-488

RS-232

Sequential

Overlapped

This command returns the source of the temperature being used for cold junction
compensation of thermocouple source and measurement.

Parameter:

Response:

where is one of the following:

INT

internal temperature sensor in use

EXT

external reference value in use

Example:

TC_REF?

INT

This example indicates that the internal sensor is in use.

TC_TYPE

IEEE-488

RS-232

Sequential

Overlapped

This command sets the Thermocouple (TC) sensor type being used for TC temperature
source and measurement. Normally, before using the TC_TYPE command to select the
TC type, use the TSENS_TYPE command to select TC mode, and afterwards set the
output temperature using the OUT command, if applicable. A change in temperature
sensors sets the output to 0 °C.

Parameters:


where is one of the following:

B

B-type thermocouple

C

C-type thermocouple

E

E-type thermocouple

J

J-type thermocouple

K

K-type thermocouple

L

L-type thermocouple

N

N-type thermocouple

R

R-type thermocouple

S

S-type thermocouple

T

T-type thermocouple

U

U-type thermocouple

XK

XK-type thermocouple

BP

BP-type thermocouple

Z

1 mV/°C

Response:

Example:

TC_TYPE J

This example sets the thermocouple type to J-type.

X

X

X

X

X

X