5

PROFIBUS SETTINGS

52

UFC 400

www.krohne.com

05/2014 - 4003118502 - AD UFC 400 PROFIBUS R02 en

M

M Electronics

Mem. Life Exceeded

The number of maximum write

cycles is exceeded for EEPROM

or FRAM or Flash (if detected).

Exchange electronics assembly.

S

S Config:

Totaliser

Tot 1 FB3 Overflow

Counter FB3 had an overflow.

After the overflow it started

again at zero.

(not supported till now)

Check totaliser format.

Tot 2 FB4 Overflow

Counter FB4 had an overflow.

After the overflow it started

again at zero.

(not supported till now)

Tot 3 FB5 Overflow

Counter FB5 had an overflow.

After the overflow it started

again at zero.

(not supported till now)

I

I Electr: Power

Failure

Tot 1 FB3 Power

Failure

Totaliser 1 invalid due to power

failure.

Check Totaliser value. Reset

Errors (A3.1).

Tot 2 FB4 Power

Failure

Totaliser 2 invalid due to power

failure.

Tot 3 FB5 Power

Failure

Totaliser 3 invalid due to power

failure.

Power Failure

Detected

Power failure. Totalisers did not

run during that time.

Reset Errors (A3.1).

I

I Electr.

Operation Info.

PROFIBUS No Data

There is no cyclic data exchange

established in connection with

this UFC 400.

PROFIBUS Baudrate

The PROFIBUS RS485 interface

is looking for a baudrate

supported.

Check PROFIBUS terminal and

connection.

PROFIBUS Passivated

PROFIBUS module is passivated. Disable PROFIBUS passivation.

PROFIBUS Clear Data

PROFIBUS Clear Data active.

Start cyclic PROFIBUS

communication.

PROFIBUS Sync Mode

PROFIBUS Sync Mode active.

Disable Sync Mode.

PROFIBUS Freeze

Mode

PROFIBUS Freeze Mode active.

Disable Freeze Mode.

Tot 1 Stopped

Tot 1 FB3 has stopped.

Select "Totaliser Function" and

choose setting other than "Stop"

(C3.1.1)

Tot 2 Stopped

Tot 2 FB4 has stopped.

Select "Totaliser Function" and

choose setting other than "Stop"

(C3.2.1)

Tot 3 Stopped

Tot 3 FB5 has stopped.

Select "Totaliser Function" and

choose setting other than "Stop"

(C3.3.1)

Tot 1 FB3 Threshold

Tot 1 FB3 Threshold exceeded.

Reset Totaliser 1 FB3 (C3.1.4).

Tot 2 FB4 Threshold

Tot 2 FB4 Threshold exceeded.

Reset Totaliser 2 FB4 (C3.2.4)

Tot 3 FB5 Threshold

Tot 3 FB5 Threshold exceeded.

Reset Totaliser 3 FB5 (C3.3.4)